Refurbished CADD Prizm VIP 6101 PCA Pump
Operator's Manual PDF Download
Prizm VIP 6101 Brochure PDF Download
The CADD-Prizm VIP Model 6101 ambulatory infusion pump has 4 delivery modes: PCA, Continuous, TPN, and Intermittent.
Dimensions: 1.7" D x 4.1" W x 5.6" H excluding cassette
Weight: 20 oz including 9 volt battery and empty 100 ml reservoir
Power: 9 volt alkaline or lithium battery
OPERATING RANGE:
PCA Delivery Mode Specifications:
Reservoir Volume: 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments
Units: ml, mg, mcg
Concentration: mg/ml, mcg/ml
Continuous Rate: 0-30 ml/hr or the mg or mcg equivalent
Demand Dose: 0-9.9 ml, Delivery rate is 125 ml/hr nominal
Demand DoseLockout: 5 min-24 hours in the following increments:
- 1 min for values between 5 and 20 minutes
- 5 minutes between 20 min and 24 hours
Max Doses Per Hour: 1-12 doses in 1 dose increments
Demand Doses Given: 0-999
Demand Dose Attmepts: 0-999
Given: 0-99999.99 in 0.01 unit increments
Clinician Bolus: 0.1 ml - 20.0 ml or mg or mcg equivalent, delivery rate is 125 ml/hr nominal
Continuous Delivery Mode Specifications:
Reservoir Volume:1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments
Continuous Rate: 0.1-350 ml/hr in the following increments:
- 0.1 for values between 0.1-100
- 1 for values between 100-350
Given: 0-99999.9 in 0.1 unit increments
TPN Delivery Mode Specifications:
Reservoir Volume: 10.0 - 9990 or Not in Use; programmable in 1 ml increments, displayed in 0.1 increments
Infusion Volume: 10-9990 ml in 10 ml increments
Infusion Period: 0 hrs 10 min - 99 hrs 40 min in 10 minute increments
Taper Down Period: 0 hrs 0 min - 99 hrs 50 min in 10 minute increments
Plateau Rate: Calculated by pump; 10-350 ml/hr
KVO Rate: Calculated by pump; 1/10 of Plateau Rate up to 5 ml/hr
Given: 0-99999 in 0.1 unit increments
Intermittent Delivery Mode Specifications:
Reservoir Volume: 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments
Dose Volume: 0.1-1000 ml in the following increments:
- 0.1 for values between 0.0 and 100
- 1 for values between 100 and 1000
Dose Duration: 1 min-24 hrs in the following increments:
- 1 min for values between 1 min and 10 min
- 5 min for values above 10 min
Duration is limited by Dose Volume so that rate does not exceed 350 ml/hr
Dose Cycle: 10 min - 96 hrs in 5 minute increments
KVO Rate: 0-10 ml/hr in 0.1 ml increments
Next Dose Start Time: 10 min - 96 hrs X min (where X equals a 10 minute increment in the 96th hour)
Given: 0-99999.9 in 0.1 unit increments