Smith Medical CADD 6500 Legacy Plus PCA Pumps - Monthly Rental
Description
Pump Resolution: 0.050 ml per pump stroke nominal
Dimensions: 1.6″ D x 3.8″ W x 4.4″ H excluding cassette
Weight: 13.8 oz including 2 AA batteries and empty 100 ml reservoir
Power Source: 2 AA alkaline batteries or AC Adapter
Maximum Infusion Pressure: 40.0 psi
System Delivery Accuracy: ±6% nominal
Maximum Time to Occlusion Alarm: 2.0 hours
Delivery Rate During Priming: Approximately 180 ml/hr. Continuous Delivery Mode Specifications
Reservoir Volume: 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments
Continuous Rate: 0.1 – 125 ml/hr; programmable in 0.1 ml/hr increments
Given: 0-99999.95 in 0.05 ml increments. Intermittent Delivery Mode Specifications
Reservoir Volume: 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments
Dose Volume: 0.1 – 1000 ml programmable in 0.1 ml increments
Dose Duration:
- 1 min to 24 hrs in the following increments
- – 1 minute for values between 1 min and 10 min
- – 5 minutes for values above 10 min
Dose Cycle: 10 min to 96 hrs in 5 min increments
KVO Rate: 0 to 125 ml/hr in 0.1 ml/hr increments
Dose Starts In:
- Immediate or 1 min to 96 hrs in the following increments
- – 1 min for values between 0 min and 10 min
- – 5 min for values between 10 min and 96 hrs
Given: 0-99999.95 in 0.05 ml increments
Dose Remaining: 1 minute increments
Pump Alarms:
- Low battery power
- Depleted battery power
- Battery dislodged
- Pump stopped
- Pump fault
- Low reservoir volume
- High delivery pressure
- Air in line
- Disposable not attached when run attempted
- Motor locked
- Upstream occlusion
- Reservoir volume empty
- Program incomplete
- Key stuck
- Disposable detached